Radioaerosol Lead Shields

radiation shielding devices for radioaerosol deliveryFor Personal and Patient Radioprotection


I.  Lead Shielding

1/16” on all surfaces

1/8” around nebulizer

1/8” on underside of cover


II. External Radiation:  With 80 mCi Tc-99m* in 2 mL in Nebulizer**:

At breathing tube opening (surface): 3.00 mr/hr
At mouthpiece opening (@1 ft.): 0.05 mr/hr
All sides & top of shield (surface): 0.05 mr/hr
Background: 0.02 mr/hr

Note:  Nebulizer was operated for 5 minutes prior to measurements to coat inner surfaces.

* 80 mCi Tc-99m is twice the recommended usage.

**Measurements made with Ludlum, Model 2, energy compensated side window G.M. survey meter.


III. Radioaerosol Lead Shield weight:

Insta/Vent™ Model IV-601 16 lbs
Aero/Vent™ MAX Model AV-501 8.2 lbs
Aero/Vent™ Jr Pole Model AV-101 8.2 lbs
Aero/Vent™ Table Top Model AV-102 13.6 lbs